Y s i – YSI ADV6600 User Manual

Page 168

Advertising
background image

Appendix B. Specifications

ADV6600

Y S I

Environmental

Page 158

Sensor Range Resolution

Accuracy

Rhodamine WT

0 to 200 µg/L as
true dye; 1,000
µg/L as dye tracer

0.1 µg/L

± 5 % of the reading or 1 ug/L,
whichever is greater

Nitrate-Nitrogen

0-200 mg/L-N

0.001 mg/L-N to
1 mg/L-N, range
dependent

± 10% of reading or 2 mg/L,
whichever is greater

Ammonium/
Ammonia-
Nitrogen

0-200 mg/L-N

0.001 mg/L-N to
1 mg/L-N, range
dependent

± 10% of reading or 2 mg/L,
whichever is greater

Chloride

0-1000 mg/L

0.001 mg/L to
1 mg/L, range
dependent

± 15% of reading or 5 mg/L,
whichever is greater

ADV Sampling Volume:

~ 0.25cc located ~ 6.8 cm from probe tip

Operating Medium:

Fresh, sea or polluted water

Operating Temperature:

-5 to +45

°C

Operating Depth:

0 to 200 feet (60 meters), sonde and all sensors except Nitrate,
Ammonia, and Chloride which are 0 to 50 feet (15 meters)

Storage Temperature:

-40 to +60

°C for sonde and all sensors except pH and pH/ORP

-20 to +60

°C for pH and pH/ORP sensors

Size:

4 in. diameter; 24.5 in. length ( 10.1cm x 62.5 cm)

Weight with Batteries and probe
guard installed:

8.8 lb. (4.0 kg)

Power:

8 Standard Alkaline C-cells, 72Wh capacity,
or External 12-15 VDC

Power Consumption:

0.92 W (continuous operation); 0.001 W (standby)

Battery Life

At least 30 days for the following conditions: 1 min averaging
interval; 15 min sample interval; 20 C; all water quality
sensors active

Computer Interface:

RS-232, SDI-12

Internal Memory:

4 Mb

Software:

IBM Compatible PC with Pentium processor or above
Windows

®

98, NT, 2000, ME, or XP

32 MB memory
5 MB available disk space
CD-ROM Drive
RS-232 Serial Port
Monitor capable of 800 x 600 resolution or better

Advertising