3-6.3. Setting Up Installed Sensors


To activate the water
quality sensors that are
installed in the
ADV6600, select 3-
Sensor
from the Sonde
Main menu. The
Sensor menu allows
you to Enable or
Disable (turn on or off)
any available sensor
and, in some cases, to
select the port in which
your sensor is installed.

Enter the corresponding number or letter to enable the sensors that are installed on your sonde. An
asterisk indicates that the sensor is enabled.

When selecting ISE3, ISE4, or any of the Optical ports, a submenu will appear. When this occurs,
make certain the sensor selected corresponds to the port in which the sensor is physically installed.

Note:
• Only ORP can be enabled as ISE2.

• ISE ports 3 and 4 can have one of three probes (nitrate, ammonium, or chloride) installed as

indicated by the submenus.

• Each optical port can have one of four probes (6026 turbidity, 6136 turbidity, chlorophyll, or

rhodamine WT) installed as indicated by the submenus.

• Only one probe of any type can be installed and activated in the ADV6600 sonde. For example,

it is not possible to use chlorophyll probes in both the Optic T and Optic C ports. However, it is
possible to use both the 6026 and the 6136 turbidity probes simultaneously

After all installed sensors have been enabled, press Esc or 0 to return to the Main Menu.

3-6.4. Setting Up Reports


The Report menu allows you to select the parameters and units of measure that will be output from
the Water Quality sensors and eventually be available for display in the ADVantage 6600 software.

Select 2-Report from the sonde Main menu. The menu below, or a similar menu, will be displayed.
The parameters listed will depend on the sensors available and enabled on your sonde.