Y s i – YSI ADV6600 User Manual

Page 72

Advertising
background image

Section 4. Calibration and Diagnostics

ADV6600

Y S I

Environmental

Page 62

Fax 937-886-9300
www.brightdyes.com


1. Accurately transfer 5.0 g of the Rhodamine WT solution into a 1000 mL volumetric flask.
2. Fill the flask to the volumetric mark with deionized or distilled water and mix well to produce a

solution that is approximately 100 mg/L of Rhodamine WT. Transfer this standard to a glass
bottle and retain it for future use.

3. Accurately transfer 5.0 mL of the solution prepared in the above step to a 1000 mL volumetric

flask.

4. Fill the flask to the volumetric mark with deionized or distilled water. Mix well to obtain a

solution, which is 0.5 mg/L in water (a 200:1dilution of the concentrated solution).

5. Store the concentrated standard solution produced above in a glass bottle in a refrigerator to retard

decomposition. The dilute standard prepared in the previous step should be used within 24 hours
of its preparation.


When rhodamine standards are required in the future, perform another dilution of the concentrated
rhodamine solution after warming it to ambient temperature. Our experience has indicated that the
concentrated solution that has been kept at cold temperatures is much more stable than the dilute
solution stored at room temperature.

It is well known that the intensity of the fluorescence of many dyes shows an inverse relationship
with temperature. This effect must also be accounted for when “calibrating” the YSI chlorophyll
sensor with rhodamine WT. Enter the calibration value from the table below corresponding to the
temperature of the standard. The table shows the approximate algal chlorophyll equivalent of 0.5
mg/L Rhodamine WT as a function of temperature.

°C

µg/L Chl to Enter

°C

µg/L Chl to Enter

30 100 18 122

28 103 16 126

26 106 14 131

24 110 12 136

22 113 10 140

20 118 8 144

Calibration


When selecting to calibrate Chlorophyll, there will be a choice of zeroing the full-scale range of the
fluorescence sensor (Fluor Zero), or calibrating with actual chlorophyll standards (µg/L 1-point, 2-
point, or 3-point). A zero calibration should always be performed before each use of the YSI 6025
sensor. Multi-point calibrations will be required infrequently. In general, it is best to perform

Advertising