Section 3. Installation and Setup

ADV6600

Y S I

Environmental

Page 32

11. Choose a Binary update.
12. If you wish to have the firmware

updated and then re-verified, be
sure the Verify new firmware
box is checked.

13. For a quicker update, check the

Use fast connection box. Note,
however, that this option should
only be utilized for hard-wired

connections between PC and sonde (i.e. not through radios).

14. Click on Start to begin the firmware update. When it is complete, the Update Firmware

window will close.


Water Quality Firmware:
1. Connect the ADV6600 to a computer.
2. Launch the ADVantage 6600 Software for Windows.
3. Choose Connect to System. Verify the COM setting and Baud Rate. Click on Connect.
4. Select Change System Settings. The Quick Setup dialog box will appear.
5. Click on Sonde Menu. A window will appear displaying the Main Menu of the 6600 firmware.
6. Press Escape, then Y. A “#” prompt will appear.
7. Type ver and press Enter. The unit will respond with the current version of 6600 firmware

installed on the board.

8. If the firmware on the board matches the current version listed on the YSI website, no upgrade

is required. If it does not, continue with Step 9.

9. Close the sonde menu and exit from ADVantage 6600.
10. Access the YSI Website (

www.ysi.com/downloads

) and fill out the Login information.

11. Locate the YSI Code Updater install file and save it to a known directory on your computer.
12. Locate the file and double click on it to install the YSI Code Updater program to your

computer.

13. Follow the onscreen instructions to complete the installation of the program files.
14. When installation is complete, click on the Start button at the bottom left of your screen.
15. Select Programs>YSI Code Updater>YSI Code Updater. The screen shown below will

appear.

Advertising