Section 11. Troubleshooting

ADV6600

Y S I

Environmental

Page 144


SYMPTOM


POSSIBLE CAUSE


ACTION


REF

Conductivity improperly
calibrated.

Follow cal procedure

Conductivity probe requires
cleaning

Follow cleaning procedure

Conductivity probe damaged

Replace probe

Calibration solution out of spec
or contaminated

Use new calibration solution

Internal failure

Return sonde for service

Conductivity
unstable or
inaccurate. Error
messages appear
during calibration.

Calibration solution or sample
does not cover entire sensor.

Immerse sensor fully.

The sensor has been disabled

Enable sensor

Water in probe connector

Dry connector; reinstall probe

Probe has been damaged

Replace the 6560 probe

Report output improperly set up Set up report output

Installed probe has
no reading

Internal failure

Return sonde for service.

Water in connector

Dry connector; reinstall probe

Temperature,
unstable or
inaccurate

Probe has been damaged

Replace the 6560 probe

Probe requires cleaning or
maintenance

Follow probe cleaning and
maintenance procedures

Probe requires calibration

Follow cal procedures

Probe has been damaged

Replace probe

Water in probe connector

Dry connector; reinstall probe

Calibration solutions out of
spec

Use new calibration solutions

Wiper is not turning or is not
synchronized.

Activate wiper. Assure rotation.
Make sure setscrew is tight.

Wiper is fouled or damaged.

Clean or replace wiper.

Turbidity probe
unstable or
inaccurate. Error
messages appear
during calibration

Internal failure.

Return probe for service.