DOsat % Local

Dissolved oxygen in % air saturation (set to 100 % at

calibration)

3-6.5. Advanced Menu Features


From the Sonde Main menu select 4-Advanced to display the sensor calibration constants,
additional setup options, and sensor coefficients/constants as shown below. The parameters listed
will depend on the sensors installed and enabled.


Select 1-Cal constants to display the calibration constants. Note that values only appear for the
enabled sensors. The following table provides the default value, operating range, and comments
relative to the calibration constants. Error messages will appear during calibration if values are
outside the indicated operating range unless the designation is “not checked”.

Parameter Default Operating

Range

Comments

Cond

5

4 to 6

Traditional cell constant

DO Gain

1

0.5 to 2.0

ORP offset mV

0.0

-100 to 100

pH offset

0.0

-400 to 400

pH gain

-5.0583

-6.07 to -4.22

NH

4

J

51.2

Not checked

NH

4

S

0.195

0.15 to 0.217

NH

4

A

1.092

Not checked

NO

3

J

99.5

Not checked

NO

3

S

-0.195

-0.217 to -0.15