5.3 Monitor

757 VA Computrace – Software

91

Sample ID [ 16 characters ; "std" ]

Identification for sample.

Sample size [ > 0 ; 10 ]

Amount of sample added to the measuring ves-
sel. The value defined on the

Determination

tab is

displayed and can be changed if desired.

Sample unit [ mL, g ; mL ]

Selection of unit for sample amount. The value
defined on the

Determination

tab is displayed and

can be changed if desired.

Cell volume (mL) [ > 0 mL ; 10 mL ]

Total volume of solution (sample + auxiliary solu-
tion, e.g. buffer) in the measuring vessel at the
start of the determination. The sample concentra-
tions Mass conc. calculated refer to this cell
volume.

Add the sample solution into the measuring ves-
sel and press

<OK>

.

START CALIBRATION

This window appears for the recording of calibra-
tion curves.

The window displays date and time of the deter-
mination start and the title of the method used.
Date and time in the format

MMDDHHMM

(month-

day-hour-minute) and the

Sample ID

are used as

default for the name of the determination file with
the recorded calibration curve to be saved auto-
matically (e.g.

02041109_std.dth

).

Calibration curve id [ 16 characters ; "std" ]

Identification for calibration curve.

Add the calibration solution into the measuring
vessel and press

<OK>

.