7.9 Diagnostic procedures

757 VA Computrace – Software

159

Perform a GLP test

The requirements of GLP (Good Laboratory Practice) include the
periodic testing of analytical instruments with regard to repro-
ducibility and accuracy using

standard operating procedures.

Metrohm suggests the procedure described below as the stan-
dard operating procedure for testing the 757 VA Computrace
Stand:

1. Click on

or

MAIN WINDOW

/ Mode / Determination

.

2. Click on

or

MAIN WINDOW

/ Window / Working method

specification

to open the

WORKING METHOD SPECIFICATIONS

window.

3. Perform the linearity test with the dummy cell (see How to per-

form a linearity test).

4. Perform the peak test with the dummy cell (see How to per-

form a peak test).

5. Click on

or

MAIN WINDOW

/ File / Load method

.

6. Select the method file

Test Pb in standard solution.mth

in the

OPEN

window and click

<OK>

. The method is loaded into the

WORKING METHOD SPECIFICATIONS

window.

7. Add 20 mL ultrapure water to the empty measuring vessel at

the 757 VA Computrace Stand.

8. Add 0.5 mL potassium chloride c(KCl) = 3 mol/L (Metrohm

No. 6.2308.020) to the measuring vessel.

9. Click on

or

MAIN WINDOW

/ Window / Monitor

to open the

MONITOR

window.

10. Start the measurement by clicking the

icon in the

MAIN

WINDOW

or the

button in the

MONITORING

window.

The

PLACE SAMPLE

window appears.

11. Add 100 µL Pb ion standard solution

ρ

(Pb) = 1 g/L (Metrohm

No. 6.2301.100) into the measuring vessel and click the

<OK>

button.

12. The sample solution is measured three times. Then the

MAN-

UAL ADDITION

window appears.

13. Use a pipette to add 100 µL Pb ion standard solution

ρ

(Pb) =

1 g/L (Metrohm No. 6.2301.100) into the measuring vessel
and click the

<OK>

button.

14. The sample solution spiked with standard addition solution is

measured three times. Then the

MANUAL ADDITION

window

appears.

Advertising