Stand-by potential – Metrohm 757 VA Computrace User Manual

Page 58

Advertising
background image

3 General settings for exploratory and determination mode

757 VA Computrace – Software

48

• During the

equilibration time

, the start potential of the

sweep is applied to the electrode.

Cleaning potential (V)

[ -5...+5 V ; -0.1 V ]

Voltage applied to the electrodes during the

Cleaning time

.

Cleaning time (s)

[ 0...80600 s ; 0 s ]

Time during which the

Cleaning

potential

is applied to the

electrodes.

Deposition potential (V)

[ -5...+5 V ; -0.9 V ]

Voltage applied to the electrodes during the

Deposition time

.

Deposition time (s)

[ 0...80600 s ; 60 s ]

Time during which the

Deposition

potential

is applied to the

electrodes.

Equilibration time (s) [ 0...80600 s ; 5 s ]

Waiting time before starting the sweep with stirrer switched
off and start potential applied to the electrodes.

Stand-by potential

The

stand-by potential can be applied to the electrodes at the

end of the measurement. It remains in force until it is switched off
manually in the

COMPUTRACE CONTROL

window or until a new

voltage is applied to the electrodes in the next measurement.

Cell off after measurement

[ on, off ; on ]

Enable/disable the switching off of the voltage applied to the
electrodes after measurement.

Stand-by potential (V)

[ -5...+5 V ; -0.1 V ]

Voltage to be applied to the electrodes after measurement if
the

Cell off after measurement

box is set to

off

.