5.1 Determination mode overview

757 VA Computrace – Software

71

5 Determination

mode

5.1

Determination mode overview

Determination mode features

The program part "Determination" is used for

quantitative volt-

ammetric analysis of inorganic and organic substances. It
comprises six different measurement modes and the possibility
for stripping techniques. Quantitative evaluation can be per-
formed via standard addition or calibration curve.
The

peak evaluation is automatic, various functions (linear, poly-

nomial, exponential) can be selected for the baseline approxi-
mation. In case of asymmetric peaks, there is a possibility to
evaluate only the front or rear peak half.
Peak evaluation and result calculation are documented in an in-
dividually compilable

report which can also contain voltammo-

grams and calibration curves.
The parameters of the voltammetric analysis are stored in a
method file. The method loaded into the working memory is the
working method which is used for performing new determina-
tions. The

determination method is the method which was used

for the recording of the loaded determination and which is stored
together with the measurement data in a determination file.

Determination mode selection

MAIN WINDOW

/ Mode / Determination

Switching to the determination mode for re-
cording and displaying of determinations.

Determination mode windows

MAIN WINDOW / Window / Working method specification (F6)

The

WORKING METHOD SPECIFICATIONS

window

will be opened or (if it is already open) closed. It
contains the specifications of the method loaded
into the working memory.