4.2 Exploratory specification window

757 VA Computrace – Software

59

Performing exploratory measurements

Measurements in the exploratory mode can be performed using
the following icons (in the

MAIN WINDOW

) or buttons (in the

EX-

PLORATORY SPECIFICATION

window):

Start measurement
The operation sequence (see section 3.4) de-
fined in the

EXPLORATORY SPECIFICATION

win-

dow is started. Each step of the operation se-
quence is listed in the first line of the status win-
dow beside the

<Start>

button.

For the running voltage sweep there is a live dis-
play in the

EXPLORATORY CURVES

window with

automatic scaling of the axes. Manual rescaling
can be done by pressing the <F4> button.

The red light at the left side of the

<Stop>

button

indicates a current overload. In this case, stop
the measurement and change the measurement
parameters.

In the second line of the status window com-
ments or error messages concerning the running
measurement are displayed.

A running measurement can be stopped, inter-
rupted and continued. Each step in the operation
sequence can be abbreviated by clicking the

<Next>

button.

Stop measurement
Stop running measurement immediately.

Hold measurement
Interrupt running measurement.

Continue measurement
Continue an interrupted measurement.

Next step
Go to next step in operation sequence.

Advertising