2 va measurement modes, Dp – differential pulse – Metrohm 757 VA Computrace User Manual

Page 41

Advertising
background image

3.2

VA measurement modes

757 VA Computrace – Software

31

3.2

VA measurement modes

DP – Differential Pulse

General:
DP
or Differential Pulse voltammetry is the most universal and
frequently used voltammetric measurement mode. It is equally
well suited for irreversible and reversible systems and offers a
high sensitivity. The DP measurement mode can be set for the
exploratory and determination mode by selecting

DP - Differential

pulse

for the

Mode

parameter in the

EXPLORATORY SPECIFICATION

or

WORKING METHOD SPECIFICATIONS

window.

Description:
For DP voltammetry, rectangular pulses with a constant amplitude
are superimposed on a stepwise rising direct voltage ramp. The
current

i is measured as a function of the voltage U immediately

before the pulse and at the end of the pulse. From the differences
between the two current measurements, peak-shaped curves are
obtained which are evaluated using linear, polynomial or expo-
nential baselines.

U

i

U

Start potential

Pulse

amplitude

Voltage step

Pulse

time

Voltage step time

Peak voltage

Peak

height

t

i = i2 - i1

t (i1) t (i2)

t (i1) t (i2)