5 Determination mode

757 VA Computrace – Software

122

a = Y.reg/offset Intercept of calibration curve
b = Slope

Slope of calibration curve in the
linear region

d = Nonlin. Non-linearity

factor

Calibration solutions

Y.offset

EV

0

0

Slope

c(eff)

Linear

Linear (trough Zero)

Nonlinear (trough Zero)

Nonlinear


3.

Measurement of sample solution
The sample solution with the unknown mass concentration
c(s) of the sample is measured one or more times (defined
by

No. of replications

). This gives:

EV(s)

Evaluation quantity of a single measure-
ment for the sample

mean(s)

Mean value of all evaluation quantities
for the sample

Std.dev.(s) Standard deviation of the individual value

EV(s) = s(s)


4.

Calculation of mass concentration c(s)
The sought mass concentration

c(s) of the sample is calcu-

lated by inserting mean(s) in the calibration function deter-
mined earlier:

mean(s) = d c(s)

4

+ b c(s) – a

EV

Sample

c(s)

mean(s)

c(eff)