4 cleaning procedure, Cleaning procedure selection, Cleaning procedure window – Metrohm 757 VA Computrace User Manual

Page 140

Advertising
background image

6 Manual control

757 VA Computrace – Software

130

6.4 Cleaning

procedure

Cleaning procedure selection

MAIN WINDOW

/ Utility / Cleaning procedure

Start cleaning procedure for solid state elec-
trodes at the 757 VA Computrace Stand.

Cleaning procedure window

The

CLEANING PROCEDURE

window serves for electrochemical

cleaning of solid state electrodes at the 757 VA Computrace
Stand.

Cleaning solution [ XXX characters ; ]

Name of cleaning solution used for electrochemical clean-
ing of solid state electrodes.

Stirrer/RDE (rpm)

[ 0...3000 rpm ; 2000 rpm ]

Revolutions per minute of the rotating disk electrode. The
stirring of the RDE remains active during all preparation pro-
cedure steps until the start of the cleaning sweep.

Purge time (s)

[ 0...80600 s ; 300 s ]

Time of inert gas purging before the first measurement of
the sample solution.

Conditioning

cycles

Electrochemical regeneration of the solid state electrode by
a freely selectable number of conditioning cycles. For every
cycle, the voltage is changed at a sweep rate of 1 V/s to the

End potential

and then decreased at the same rate back to

the

Start potential

.

Advertising