Standard addition peaks displaced, No addition, Spikes / signal jump in voltammogram – Metrohm 757 VA Computrace User Manual

Page 182

Advertising
background image

8 Troubleshooting

757 VA Computrace – Software

172

Standard addition peaks displaced

With

all types of electrodes:

1. Standard solutions have been made too acidic.
2. Buffering capacity of the electrolyte is not sufficient: increase

electrolyte volume.

3. Electrolyte solution too old: make up a new one. Its working

life with organic additives may be as short as 1 day or less.

With

HMDE:

1. If HMDE is used potential displacements of more than 20…30

mV are often normal and have to be accepted; particularly in
adsorption voltammetry.

With

RDE/SSE:

1. Electrode surface overcharged: reduce sample volume.

No addition

With

all types of electrodes:

1. Has the correct standard solution been used or is the concen-

tration of the solution too low: increase the volume of the
standard addition solution or use a higher concentration or
reduce the sample amount accordingly.

2. If organic components are present: carry out a UV digestion

or other suitable sample preparation.

3. Concentration of the analyte is too high: dilute.
4. Electrolyte solution too old: make up a new one. Its working

life with organic additives may be as short as 1 day or less.

With

HMDE:

1. Addition solution with metal complexing solution (time reac-

tion).

Spikes / signal jump in voltammogram

1. For MME: check the electrode.
2. Reduce the dynamic range of the potentiostat (see Potentio-

stat, section 3.3).