Dc – sampled direct current – Metrohm 757 VA Computrace User Manual

Page 45

Advertising
background image

3.2

VA measurement modes

757 VA Computrace – Software

35

DC – Sampled Direct Current

General:
DC
or Sampled Direct Current voltammetry is the classic,
simplest voltammetric measurement mode with limited sensitivity.
It is mainly used for the investigation of reversible redox systems.
The DC measurement mode can be set for the exploratory and
determination mode by selecting

DC - Sampled direct current

for

the

Mode

parameter in the

EXPLORATORY SPECIFICATION

or

WORKING METHOD SPECIFICATIONS

window.

Description:
For DC voltammetry, the direct voltage applied to the working
electrode is continuously changed and the resultant current

i

which flows measured as a function of the voltage

U. For DME

and SMDE this normally provides wave-shaped curves which can
be evaluated in the exploratory mode using the tangent method.

U

t

t (i)

Start potential

Voltage step

Voltage step time

Half-wave voltage

Wave current

i

U

t (i)

Advertising