Metrohm 757 VA Computrace User Manual

Page 90

Advertising
background image

5 Determination mode

757 VA Computrace – Software

80

Nonlinear Regression (through Zero)

The regression is calculated with a nonlinear cur-
ve forced to the zero point. This option is only
available if

Record calibration curve

is selected for

Calibration

.

Peak evaluation [ Height, Area, Derivative ; Height ]

Selection of peak evaluation quantity:

Height

Peak height from baseline to peak maximum.

Area

Peak area between peak curve and calculated
baseline.

Derivative

Difference between positive an negative maxi-
mum of the first derivative of the voltammogram.

Minimum peak width (V.step) [ 5 ; 5 ]

Minimum peak width for peak recognition by
number of

Voltage steps

(= number of measure-

ment points).

Minimum peak height (A) [ > 50 pA ; 100 pA ]

Minimum peak height for peak recognition.

Smooth factor [ 1...6 ; 4 ]

Smoothing power of the Savitzky/Golay smooth-
ing of the baseline (

1

= minimum smoothing,

6

=

maximum smoothing).

Reverse peaks [ on, off ; off ]

Enable peak evaluation of reverse peaks (peaks
with opposite direction compared to the sweep
or pulse direction).

Reverse sweep [ on, off ; off ]

Enable peak evaluation of the reverse sweep of
cyclic sweeps (only available with CV).

Advertising