Metrohm 757 VA Computrace User Manual

Page 139

Advertising
background image

6.3 Film deposition

757 VA Computrace – Software

129

Purge time (s)

[ 0...80600 s ; 300 s ]

Time of inert gas purging before the first measurement of
the sample solution.

Conditioning

cycles

Before deposition, the solid state electrode can be electro-
chemically regenerated by a freely selectable number of
conditioning cycles. For every cycle, the voltage is changed
at a sweep rate of 1 V/s to the

End potential

and then de-

creased at the same rate back to the

Start potential

.

Start potential (V)

[ -5...+5 V ; -1.2 V ]

Start voltage for the cyclic conditioning sweep.

End potential (V)

[ -5...+5 V ; -0.1 V ]

Final voltage for the cyclic conditioning sweep.

No. of cycles

[ 0...X ; 0 ]

Number of conditioning cycles.

Deposition potential (V)

[ -5...+5 V ; -0.9 V ]

Voltage applied to the electrodes during the

Deposition time

.

Deposition time (s)

[ 0...80600 s ; 60 s ]

Time during which the

Deposition

potential

is applied to the

electrodes.

Cleaning potential (V)

[ -5...+5 V ; -0.1 V ]

Voltage applied to the electrodes during the

Cleaning time

.

Cleaning time (s)

[ 0...80600 s ; 0 s ]

Time during which the

Cleaning

potential

is applied to the

electrodes.

Sweep

Parameters of DP sweep used at the end of the film deposi-
tion for checking the electrode, see VA measurement
modse
, section 3.2.

Cell off after measurement

[ on, off ; on ]

Enable/disable the switching off of the voltage applied to the
electrodes after measurement.

Stand-by potential (V)

[ -5...+5 V ; -0.1 V ]

Voltage to be applied to the electrodes after measurement if
the

Cell off after measurement

box is set to

off

.








Advertising