3 General settings for exploratory and determination mode

757 VA Computrace – Software

40

Sweep parameters:

Hydrodynamic (measurement) [ on, off ; off ]

Enable/disable stirring of the RDE/SSE during the sweep.

Start potential (V)

[ -5...+5 V ; -0.9 V ]

Start voltage for the voltage sweep.

End potential (V)

[ -5...+5 V ; -0.1 V ]

Final voltage for the voltage sweep.

Voltage step (V)

[ > 0 V ; 0.006 V ]

Voltage step for direct voltage ramp.

Sweep rate (V/s)

[ > 0 V/s ; 0.1 V/s ]

Ramp slope =

Voltage step / Voltage step time

.

No. of sweeps

[ > 0 ; 1 ]

Number of cyclic sweeps to be performed.

Save last ... sweeps]

Number of cycles to be saved.

Notes:
• The CV measurement mode can only be used with HMDE or

RDE electrodes.

• The following condition applies to the definition of

Voltage step

and

Sweep rate

:

Voltage step time = Voltage step / Sweep rate > 270 µs