Curves with high noise – Metrohm 757 VA Computrace User Manual

Page 178

Advertising
background image

8 Troubleshooting

757 VA Computrace – Software

168

4. Has the electrode been conditioned (e.g. by using

Condition-

ing cycles

and

Cleaning potential

) ?

5. If the concentration to be determined is considerably higher

than expected: reduce sample volume.

6. The background current is normally higher if RDE is used in

place of MME; a background current of several 100 nA is
possible.

Curves with high noise

With

all MME types:

1. Have stirring or degassing been switched off during the

measurement?

2. Check contact between needle and capillary. If necessary,

clean the MME (see Hardware Manual).

3. Check tapping mechanism on stand. If tapping strength is too

weak, turn corresponding slotted screw on the valve block
during operation in an anticlockwise direction until a drop falls
each time the tapper is triggered (see Hardware Manual).

4. Electrolyte solution too old: make up a new one. Its working

life with organic additives may be as short as 1 day or less.

With

DME/SMDE:

1. If the electrode drops irregularly: check the MME.

Adjust sealing needle. If necessary, change capillary or re-
place sealing needle (see Hardware Manual).

2. If the electrode drops much too quickly: reduce the

Voltage

step time

on the

Voltammetric

tab of the

EDIT WORKING

METHOD PARAMETERS

window.


With

HMDE:

1. If the electrode surface is overcharged: check deposition po-

tential and time.

2. If no drops are at the capillary: change the capillary or replace

sealing needle (see Hardware Manual).