2 Main window

757 VA Computrace – Software

24

Note:

Method 2

must be set on the 813 Compact

Autosampler.

Time to move to sample (s) [ 30 s ; 25 ... 2147483647 s ]

Maximum time needed for the 813 Compact
Autosampler to move from one sample to the
other. (minimum time allowed:

25 s

).

Sample transfer time (s) [ 300 s ; 30 ... 2147483647 s ]

Time to transfer the sample solution from the
sample vessel to the measurement vessel using
the peristaltic pump of the 813 Compact Auto-
sampler (minimum time allowed:

30 s

).

No. of rinsing cycles [ 3 ; 0 ... 2147483647 ]

Number of rinsing cycles.

Siphoning time (s) [ 25 s ; 0 ... 2147483647 s ]

For each rinsing cycle, the measuring vessel is
siphoned off during this time using a 772 Pump
Unit.

Rinsing time (s) [ 5 s ; 0 ... 2147483647 s ]

For each rinsing cycle, the measuring vessel is
rinsed during this time using a 772 Pump Unit.

Test the 813 Compact Autosampler with the set
automation parameters.

Note: Before starting the test, switch on the 813 Compact Auto-
sampler, set

Method 2

at the Autosampler and place two sample

vessels filled with water on the sample rack.

For details on the use of the 813 Compact Autosampler, see 813
Instructions for Use
.



Save settings

MAIN WINDOW / Settings / Save now

This function saves the actual settings of the
software: Open windows, window position and
size, general settings.

MAIN WINDOW / Settings / Save on exit

If this function is enabled, the software settings
are stored when the software is quit.