5.2 Working method specifications

757 VA Computrace – Software

87

Dosimats

The automatic use of 665 or 765 Dosimats for the addition of so-
lutions has to be defined in the

DOSIMATS

window which is

opened by clicking the

button in

WORKING

METHOD SPECIFICATIONS

window.

Note: Make sure, that the desired Dosimat has also been
checked on the

Hardware

tab of the

GENERAL SETTINGS

window

(see section 2.7).

Dosimat used [ on, off ; off ]

Checkbox for Dosimats used for automatic addi-
tion of solutions. For the addition of standard so-
lutions, the Dosimat must have the same number
as the

Standard solution No.

entered on the

Sub-

stances

tab of the

EDIT WORKING METHOD

PARAMETERS

window.

Use for predose [ on, off ; off ]

Use of a Dosimat for addition of an auxiliary solu-
tion (e.g. buffer) at the start of the determination.

Volume of predose (mL) [ > 0.01 mL ; 0 mL ]

Volume of the auxiliary solution (e.g. buffer) to be
added at the start of the determination.

Content [ 46 characters ; "solution" ]

Remarks regarding the solution.