7 How to ...?

757 VA Computrace – Software

142

5. Place the analysis solution in the measuring vessel at the 757

VA Computrace Stand.

6. Click on

or

MAIN WINDOW

/ Window / Monitor

to open the

MONITOR

window.

7. Start the measurement by clicking the

icon in the

MAIN

WINDOW

or the

button in the

MONITORING

window.

8. Follow the instructions in the appearing message windows.

Perform a test determination with the Pb test method

With the aid of this example method for the determination of lead
in the ion standard solution supplied using the DME, you can eas-
ily check whether the 757 VA Computrace System is functioning
properly.

1. Click on

or

MAIN WINDOW

/ Mode / Determination

.

2. Click on

or

MAIN WINDOW

/ Window / Working method

specification

to open the

WORKING METHOD SPECIFICATIONS

window.

3. Click on

or

MAIN WINDOW

/ File / Load method

.

4. Select the method file

Test Pb in standard solution.mth

in the

OPEN

window and click

<OK>

. The method is loaded into the

WORKING METHOD SPECIFICATIONS

window.

5. Add 20 mL ultrapure water to the empty measuring vessel at

the 757 VA Computrace Stand.

6. Add 0.5 mL potassium chloride c(KCl) = 3 mol/L (Metrohm

No. 6.2308.020) to the measuring vessel.

7. Click on

or

MAIN WINDOW

/ Window / Monitor

to open the

MONITOR

window.

8. Start the measurement by clicking the

icon in the

MAIN

WINDOW

or the

button in the

MONITORING

window.

The

PLACE SAMPLE

window appears.

9. Use a pipette to add 100 µL Pb ion standard solution

ρ

(Pb) =

1 g/L (Metrohm No. 6.2301.100) into the measuring vessel
and click the

<OK>

button.

10. The sample solution is measured three times. Then the

MAN-

UAL ADDITION

window appears.

11. Use a pipette to add 100 µLl Pb ion standard solution

ρ

(Pb) =

1 g/Ll (Metrohm No. 6.2301.100) into the measuring vessel
and click the

<OK>

button.