Metrohm 757 VA Computrace User Manual

Page 159

Advertising
background image

7.6 Standard addition technique

757 VA Computrace – Software

149

13. If the standard addition should be done with variable addition

volumes, click the button in the

Volume

column to open

the

EDIT VARIED ADDITION

window, enter the variable addition

volumes in the

Addition

fields, and close this window by click-

ing

<OK>

.

14. Close the

EDIT WORKING METHOD PARAMETERS

window by

clicking

<OK>

.

15. Place the sample solution in the measuring vessel at the 757

VA Computrace Stand.

16. Click on

or

MAIN WINDOW

/ Window / Monitor

to open the

MONITOR

window.

17. Start the measurement by clicking the

icon in the

MAIN

WINDOW

or the

button in the

MONITORING

window.

18. Enter the

Sample ID

(used as part of the determination file

name) in the

PLACE SAMPLE

window and click

<OK>

.