8 Troubleshooting

757 VA Computrace – Software

170

No peak found

With

all types of electrodes:

1. The peak is only displaced: adjust the half-wave potential and

recalculate the results.

2. The sample concentration is too low: increase the sample

volume or the amount of sample.

3. Are the

Start potential

and

End potential

correct?

4. Electrolyte solution too old: make up a new one. Its working

life with organic additives may be as short as 1 day or less.

5. Are organic components present? Carry out a UV digestion or

other suitable sample preparation.


With

DME/SMDE:

1. The concentration of the ion to be determined is too low: use

HMDE (inverse voltammetry) instead of DME or SMDE.


With

HMDE:

1. Has the complex former been forgotten? (adsorptive voltam-

metry).

2. The

Deposition time

in the inverse voltammetry is too short: in-

crease the time on the

Voltammetric

tab of the

EDIT WORKING

METHOD PARAMETERS

window.

3. No Hg drops at the capillary: check MME. Adjust sealing

needle. If necessary, change capillary or replace sealing nee-
dle (see Hardware Manual).


With

RDE/SSE:

1. The background current is too high: repolish the electrode.
2. The

Deposition time

in the inverse voltammetry is too short: in-

crease the time on the

Voltammetric

tab of the

EDIT WORKING

METHOD PARAMETERS

window.

Peak is in the highest µA range

With

all types of electrodes:

1. The concentration of the ion to be determined is too high: re-

duce the sample volume and carry out the analysis again.

With

HMDE:

1. The

Deposition time

is too long: Reduce the time.

2. If necessary use a SMDE or DME electrode instead of HMDE.