Record calibration curve automatically – Metrohm 757 VA Computrace User Manual

Page 162

Advertising
background image

7 How to ...?

757 VA Computrace – Software

152

13. Start the measurement by clicking the

icon in the

MAIN

WINDOW

or the

button in the

MONITORING

window.

14. Enter the

Calibration curve id

(used as part of the determina-

tion file name) in the

START CALIBRATION

window and click

<OK>

.

15. Each time a solution exchange is required in the

BATCH SO-

LUTION EXCHANGE

window, replace the solution measured by

the next calibration solution.

Record calibration curve automatically

This method is used for preparing different calibration solutions
by adding automatically several times a concentrated standard
solution to the measuring solution using 665 or 765 Dosimats.
Proceed as follows:
1. Install Dosimats to the 757 VA Computrace Stand (see How to

install Dosimats).

2. Click on

or

MAIN WINDOW

/ Mode / Determination

.

3. Click on

or

MAIN WINDOW

/ Window / Working method

specification

to open the

WORKING METHOD SPECIFICATIONS

window.

4. Load the desired method into the

WORKING METHOD SPECIFI-

CATIONS

window (see How to load a method).

5. Select

Record calibration curve

in the

Calibration

field of the

WORKING METHOD SPECIFICATIONS

window.

6. Select

Automatic

in the

Addition

field of the

WORKING METHOD

SPECIFICATIONS

window.

7. Click

to open the

EDIT WORKING METHOD PA-

RAMETERS

window.

8. Select the

Determination

tab (see section 5.2) and enter the

Cell volume

and the number of additions in the

No. of additions

field.

9. Select the

Substances

tab (see section 5.2) and make sure

that for each substance entered in the table the number of the
single or mixed standard solution, its concentration and its
volume is defined. The

No.

of the standard solution must be

identical to the number of the Dosimat used for automatic
addition of this solution.

10. If the addition of the standard solution should be done with

variable addition volumes, click the button in the

Volume

column to open the

EDIT VARIED ADDITION

window, enter the