8 Troubleshooting

757 VA Computrace – Software

162

8.3 Software

problems

Error message "No access to software"

If the login fails because no password is known any more, pro-
ceed as follows:
1. Deinstall the software (see Deinstallation, section 1.3).
2. Reinstall the software (see Software installation, section 1.3).

Error message "The file 'ecousb.sys' is needed"

This message indicated problems with the USB connection. Pro-
ceed as follows:
1. Insert the installation CD into the CD drive.
2. Click on

<Browse>

. Select the CD drive and click on

<OK>

.

Wrong language in Help

If you want to change the help language, reinstall the software us-
ing the

Modify

option and select the desired language.



8.4 Dosimat

problems

Dosimat does not work

1. Check if the Dosimat is switched on.
2. Check the connecting cable between the Dosimat and 757 VA

Computrace Stand.

3. Click on

MAIN WINDOW / Settings / General settings

and check

the entries on the

Hardware

tab of the

GENERAL SETTINGS

window (see Hardware settings for Dosimats, section 1.3).

4. Exit/restart the 757 VA Computrace software.
5. Activate the Dosimat in the

DOSIMAT

window for every method

used (see Dosimats, section 5.2).

6. Check if a solution number

No.

is defined on the

Substances

tab of the

EDIT WORKING METHOD PARAMETERS

window.

Note: If Dosimats are connected to the 757 VA Computrace
Stand, the instruments must always be switched on in the se-
quence Dosimat → 757 → PC.