5 Determination mode

757 VA Computrace – Software

78

Substances

The

Substances

tab of the

EDIT WORKING METHOD PARAMETERS

window contains parameters for the definition and recognition of
substances, for the definition of standard solutions, for peak
evaluation and results calculation. The parameters displayed de-
pend on the calibration technique selected in the

WORKING

METHOD SPECIFICATIONS

window.

Substance [ 8 characters ; ]

Substance name. For the assignment of a found
peak to this substance the checkbox on the left
side of the substance name must be checked.

Peak pos. +/- (V) [ -5...+5 V ; 0 V ]

Position of the peak voltage for the substance
and tolerance for this verification voltage.

Bsln.

Parameters for baseline evaluation (details see
Baseline). Click the button to open the

BASELINE

window for the selected substance.

Standard

solution

Definition of addition solutions for standard addi-
tion or recording of calibration curves. These pa-