Metrohm 757 VA Computrace User Manual

Page 136

Advertising
background image

6 Manual control

757 VA Computrace – Software

126

10 mA ... 100 nA

Selection of current range for measurement in
the manual control mode.

I ovl

Indication of current overload by red light.

Switch on/off current measurement. If switched
on, the set

Potential

is applied to the electrodes

and the current is measured continuously. This
mode is indicated by the red light beside the

<Cell>

button.

Potential (V) [ -5...+5 V ; 0 V ]

Voltage to be applied to the electrodes.

Potential (V) [ read only ]

Display of current voltage applied to the elec-
trodes.

Current (xA) [ read only ]

Display of measured current.

No electrode

No electrode connected to the 757 VA Com-
putrace Stand. This setting is useful for changing
the electrode at the stand.

DME

Selection of the Dropping Mercury Electrode
(DME).

SMDE

Selection of the Static Mercury Drop Electrode
(SMDE).

HMDE

Selection of the Hanging Mercury Drop Electrode
(HMDE).

RDE/SSE

Selection of the Rotating Disk electrode (RDE).

Switch on/off inert gas purging.

Switch on/off stirrer/RDE with the set

RDE/Stirrrer

speed

.

RDE/Stirrer speed (rpm) [ 0...3000 rpm ; 2000 rpm ]

Revolutions per minute of the stirrer/RDE.

DME: Switch on free dropping at the MME.

SMDE: Dropping in intervals of ca. 1 s at the
MME.
HMDE: Formation of a new single mercury drop
at the MME .

Drop size [ 1...9 ; 4 ]

Size of the mercury drop (surface 0.15 mm

2

...

0.60 mm

2

).

Advertising