Zooming – Metrohm 757 VA Computrace User Manual

Page 109

Advertising
background image

5.4 Determination curves

757 VA Computrace – Software

99

Automatic

Switch on/off the automatic peak baseline
evaluation.

Begin

Manual setting of the start base point for baseline
evaluation. The base point can be moved either
by manually changing the voltage value in the
first field or by clicking the buttons of the
second field indicating the current value. This
field can only be edited if

Automatic

is disabled.

End

Manual setting of the end base point for baseline
evaluation. The base point can be moved either
by manually changing the voltage value in the
first field or by clicking the buttons of the
second field indicating the current value. This
field can only be edited if

Automatic

is disabled.

Type [ Linear, Polynomial, Exponential ; Linear ]

Selection of the baseline type. This field can only
be edited if

Automatic

is disabled.

Scope [ Whole peak, Front end, Rear end ; Whole peak ]

Selection of the range for baseline evaluation.
This field can only be edited if the

Linear

baseline

type is selected.

Start baseline evaluation with the current parame-
ters entered in the

EDIT BASELINE

window.

Zooming

Curve regions in the determination curves subwindow can be
enlarged by zooming the desired area while pressing the left
mouse button ("drag a box"; reset see Auto scaling).