2 fast access to the calibration, 3 exiting the programming and calibration – Flintec FT-10xx User Manual

Page 23

Advertising
background image

FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

5.2 Fast Access to the Calibration

The instrument has fast access calibration feature to earn time to the service technician. If only the
calibration adjustment is needed, follow the steps below to access the calibration parameters fast.

Display

Operation

[123.456 kg ]

Press

key until [ PASSWr ] prompts seen.

[PASSWr]

Press

+

+

keys sequentially.

[ --- ]

Press

key for confirm.

[ 310 ]

Zero Adjustment parameter.

“Calibration”

Press

key to start zero adjustment.

Or press

key to access span calibration without zero adjustment.

5.3 Exiting the Programming and Calibration

If you press

key on which parameter you are, you will get out of the active sub-block and reach the

next sub-block. If you press

key again, you will get out of the active block and reach the next main

block. If you press

key once again, the [ SAvE ] message appears on the display. Here you can press

key to save the changes into the memory, or you can press

key to store the changes until the

power goes off, or you can press

key to abort changes. [

Waıt ] message will be seen on the display

for a little while, and automatically get back to weighing mode.
Especially for legal metrological usage, please don’t forget to turn the power off and “OFF” position the
calibration DIP switch to start operation.