FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

5.4.4

Entries [14-]

In this block you can enter the initial CN.

[142 ] Label No ( CN )

[ XXXXXX ]
The desired value is entered via < Tare > and < Zero > keys and saved by pressing < Enter > key.
If the number exceeds 65535, it will automatically reset and begin from 1 again.

5.4.5 Scale Block [2--]

[20-] Set Up

The parameters about weighing operation are being entered here.

[200 0] Approved

0 : No

1 : OIML

Warning:

Scale type cannot be programmed to force mode if OIML is selected.

[201 0] Increased Indication

0 : by pressing key

1 : Always increased indication

[202 0] Power On Zero

According to the selection below, during power on, if the weight is in the percentage of the capacity,
the scale will automatically be zeroed. If the weight is not in zeroing range, the display will show
[E E E ] message until pressing < Enter > key.
0 : Disable

1 : ± 2%

2 : ± 10%

[203 3] Zeroing Range

0 : Disable

1 : ± 2%

2 : ± 20%

3 : ±50%

[204 0] Auto Zero Tracking

AZT automatically readjusts the scale to zero for compensating selected small deviation per second
around centre of zero.
0 : Disable

1 : ± 0,5e

2 : ± 1e

3 : ± 3e

[205 1] Tare

0 : Disabled
1 : Multi tare via key
2 : Tare via key if scale is in gross mode

[206 2] Motion Detector

This parameter defines the sensitivity level which will determine what is considered as stable.
0 : ± 0,3e

1 : ± 0,5e

2 : ± 1e

3 : ± 2e

4 : Disable

[207 0.0] Stability Period

If the scale is stabile during this time, the scale is accepted as a stabile to process zeroing, tare,
print etc. commands. It can be entered up to 9.9 sec.