Flintec FT-10xx User Manual

Page 50

Advertising
background image

FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

0

None

1

Zero

2

Tare

3

Clear

4

Print

40010

R/W

20

Not used

40030

R/W

1

Calibratio
n

Dec

Description

0

None

188

Adjust Zero Calibration

220

Adjust Span Calibration

( 1 )

236

Total Load Cell Capacity

( 1 )

eCal
Coefficients

250

Average mV/V

( 1 )

171

Dead Load

( 1 )

23205

Save the coefficients of eCal

40031

R/W

2

Span Calibration Value / LC capacity / mV value / Dead load value

40033

R

1

Calibratio
n Status


D0 .. D7

Calibratio
n Process
Status

De
c

Description

1

Ready for calibration

3

Zero calibration in process …

4

Span calibration in process …

9

Error (Refer to D8 ... D15 )

D8 .. D15

Calibratio
n Errors

1

Calibration Timeout
- Restart calibration

2

ADC Error
- Re-energize the instrument
- If seen again, change the board.

3

Instrument can not be calibrating
- Check load cell cable
- Re-energize the instrument

34

Instrument can not be calibrating
- Load cell signal is very low or too high

35

Calibration Error
- Calibration loading is not enough
- Check test weight loading
- Check load cell connections

36

Calibration load value entry Error
- Test weight is too small. Increase the test
weight

37

Scale unstable
- Wait until scale become stable
- Check grounding wiring

38

The Calibration DIP switch is not ON
position.
- Check the calibration DIP switch.

40034

R/W

37

Not used

40071

R

1

Indicated weight

40072

R

1

Status

40073

R

1

Tare weight

40074

R

1

Gross weight

40100

R

1

Voltage of
Power Supply

Voltage of power supply is indicated with 0.1 V
increment.
For example: 23.4 VDC is indicated as integer 234
value.


40200

R

1

Status of Inputs

D0

Input-1