Example – Flintec FT-10xx User Manual

Page 46

Advertising
background image

FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

X

Read weight value in increased resolution


Command :

[ADR][X]

Response :

[ADR][X][STATUS][SIGN][WEIGHT VALUE]

Example

:

Command : 01X

Response : 01XS+00123.41 (weight is stable and 123.41) or

01XD+00123.41 (weight is dynamic and 123.41) or

01XE

(Error)

Comments :
The response includes weight data with divided the increment to 10.

Z

Zero


Command : [ADR][Z]
Response : [ADR][Z][A]

(Zeroed)

[ADR][Z][N]

(Zeroing could not be operated)

[ADR][Z][X]

(Zeroing is disabled)

Comments :
Zero command can not work in net weighing.
Weight must be in zeroing range for all operating modes.
Status must be stable in 2 seconds time out delay. If so, Ack is send.
If it can not be stable in time out delay, Nack is send.

Checksum Calculation:
CHK is transmitted as two ASCII characters calculated with the Checksum formulation.
Checksum = 0

– (SUM of all response data before CHK)

Example:

Read stable current weight data.


BSI Examples: (CHK is enabled and instrument address is 01)

Command: 01P[CHK][CR][LF]

Checksum

= 0 − (0x30 + 0x31 + 0x50)

= 0 − 0XB1

= 0x4F

= Char ‘4’ and ‘F’



Response: 01PS+000123.4[CHK][CR][LF]

Checksum

= 0 − (0x30 + 0x31 + 0x50 + 0x53 + 0x2B + 0x30 + 0x30 + 0x30 + 0x31+ 0x32 + 0x33 + 0x2E +

0x34)

= 0 − 0x02B7

= 0x49

= Char ‘4’ and Char ‘9’