FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

Calibration Status (always 32 bit integer)
1

st

Dword descriptions when read command is ‘Calibration Status’. Refer to 2

nd

Dword of PLC Output to FT-

10 PB Input

Bit Number

1

st

Dword Description

D31 … D16

Not in use

D15 … D8

Calibration
Errors

Bin

Dec

Descriptions

0000 0001

1

Calibration Timeout
- Restart calibration

0000 0010

2

ADC Error
- Re-energize the instrument

0000 0011

3

Instrument can not be calibrating
- Check load cell cable
- Re-energize the instrument

0010 0010

34

Instrument can not be calibrating
- Load cell signal is very low or too high

0010 0011

35

Calibration Error
- Calibration loading is not enough
- Check test weight loading (Write test weight
value to 1

st

Dword of PLC Output to FT-10 PB

Input then restart the calibration)
- Check load cell connections

0010 0100

36

Calibration load value entry Error
- Test weight is too small. Increase the test weight

0010 0101

37

Scale unstable
- Wait until scale become stable
- Check grounding wiring

0010 0110

38

The Calibration DIP switch is not ON position.
- Check the calibration DIP switch.

D7 ... D0

Calibration
Process
Status

0000 0001

1

System ready for calibration

0000 0011

3

Zero calibration in process

0000 0100

4

Span calibration in process

0000 1001

9

Error ( Refer to Calibration Errors )

Table 9.1

– Calibration Status

PLC Output to FT-10 PB Input

Bitwise of a Dword:

Dword
(R/W)

D31 D30 D29 D28 D27 D26 D25 D24 D23 D22 D21 D20 D19 D18 D17 D16

D15 D14 D13 D12 D11 D10 D9

D8

D7

D6

D5

D4

D3

D2

D1

D0


1

st

Dword

Next Dword defines the usage of this Dword.

2

nd

Dword

Out
5

Out
4

Out
3

Out
2

Out
1

Expanded Commands List

Not in use

Command List

Read Data Selection

New