FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

10. P

ROFINET

(

ONLY

FT-10

PN

)

The Profinet interface operates at 100Mbit, full duplex, as required by Profinet. GSDML file is available in CD
which is supplied together with the instrument.

[06-] Profinet

This sub-block includes the parameters related with the Profinet interfaces of FT-10 indicator.

[060 0] Data Format

0 : Signed 32 bit integer, no decimal point implied
1 : 32 bit float, decimal point implied


There are 3 announcement LEDs on rear panel to show the instrument status in operation and 3 LEDs (MS,
NS, LINK) for showing Profinet status in operation.
The meanings of these LED’s are;

MS Module Status LED

LED State

Description

Comment

Off

Not power or not initialized

No power or profinet module is in
initialization state

Green

Initialized

Green, 1 flash

Initialized, diagnostic event(s) present

Diagnostic is active

Red

Exception error

There is an exception error

Red, 1 flash

Configuration Error

Check GSDML configuration

Red, 2 flashes

IP Address Error

IP address not set

Red, 3 flashes

Station Name Error

Station name not set

Red, 4 flashes

Internal Module Error

Re-energize the instrument.
If seen again, change the board.


LINK/Activity LED

LED State

Description

Comment

Off

No Link

No link, no communication present

Green

Link

Ethernet link established, no
communication present

Green, flickering

Activity

Ethernet link established,
communication present


NS Network Status LED

LED State

Description

Comment

Off

Not online /No power

Check power and cable

Green

On-line (RUN)

-

Green, flashing

On-line (STOP)

-

1

NS Network Status LED

2

MS Module Status LED

3

Link/Activity LED

4

Profinet interface