FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

Page 77 of 82

13. T

ROUBLE

S

HOOTING

FT-10 weighing indicator had been designed as a very reliable and virtually error free instrument. However, if
an error occurs, do not attempt to repair the equipment before understanding what caused the error. Note
the problems you have with your instrument and the error messages shown on the display. Then try to solve
the problem according to the error table given below.

Error Table

ERROR
CODE

DESCRIBTION

THINGS TO DO

Under

Weight is too low

- Check the load
- Load cell or instrument could be broken.

Over

Over Load

ADC Out

Load exceeds the operation
range

- Check the load
- Check the calibration
- Load cell or instrument could be broken.

Err 1

ADC error

- Re-energize indicator
- Call FLINTEC

Err 2

ADC error

- Re-energize indicator
- Call FLINTEC

Err 3

Indicator cannot be calibrated

- Check load cell cable and load then start
calibration again

Err 10

EEPROM error

- Configure the instrument
- EEPROM broken

Err 20

Calibration error

- Calibrate the indicator..

Err 21

Configuration error

- Configure the indicator.

Err 22

Tare, CN, Total weight and the
SP in use error

- Check SP, PT and ID entries.
- Check Tare, CN and Total weight

Err 26

Set point error

- Reload Set points.

Err 27

Indicator is not calibrated

- Calibrate the indicator

Err 30

Processor Error

- Call FLINTEC

Err 34

Indicator can not be calibrating

- Load cell signal is negative , very low or too
high

Err 35

Calibration Error

- Calibration loading is not enough.
- Check test weight loading.

Err 36

Calibration load value entry Error

- Test weight is too small. Increase the test
weight.

Err 37

Scale unstable

- Wait until scale become stable.
- Check grounding wiring.

Err 47

Main pcb info error

- Call FLINTEC

Err 61

Eeprom is not installed or broken

- Call FLINTEC

Err 70

Modbus selection error

- Check data format of other interfaces.
- Other interfaces should not be Modbus.

E XXXX

Hardware error

- Call FLINTEC

14. P

ARAMETER

-

D

EFAULT

T

ABLE

0--

Interface Block

14-

Entries

00-

RS 232C

142

Label No entrance

000 Data Format

3 = BSI

2--

Scale Block

001 Baud rate

3 = 9600

20-

Set up

003 Address

0

200

Approved

0 = No

004 Data length and Parity 0 = 8 bit, no parity

201

Increased indication

0 = x10 key

005 Checksum

0 = Disable

202

Power on zero

0 = Disable

006 Carriage Return

1 = Enable

203

Zero range

3 = ± %50

007 Line Feed

1 = Enable

204

Auto zero tracking

0 = Disable

008 Response Speed

0 = immediately

205

Tare

1 = With Tare
Key