FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

D1

Input-2

D2

Input-3

D3

Input-4

40201

R/W

1

Status of Outputs

D0

Output-1

D1

Output-2

D2

Output-3

D3

Output-4

D4

Output-5

40202

R/W

2

Setpoint 1 Low

( 2 )

40204

R/W

2

Setpoint 1 High

40206

R/W

2

Setpoint 2 Low

( 2 )

40208

R/W

2

Setpoint 2 High

40210

R/W

2

Setpoint 3 Low

( 2 )

40212

R/W

2

Setpoint 3 High

40214

R/W

2

Setpoint 4 Low

( 2 )

40216

R/W

2

Setpoint 4 High

40218

R/W

2

Setpoint 5 Low

( 2 )

40220

R/W

2

Setpoint 5 High


42000

R/W

1

Filter

Refer to parameter [120] on page 29

42001

R/W

1

Power On Zero

Refer to parameter [202] on page 30

42002

R/W

1

Zeroing Range

Refer to parameter [203] on page 30

42003

R/W

1

Auto Zero Tracking

Refer to parameter [204] on page 30

42004

R/W

1

Tare

Refer to parameter [205] on page 30

42005

R/W

1

Motion Detector

Refer to parameter [206] on page 30

42006

R/W

1

Stability Period

Refer to parameter [207] on page 30

42007

R/W

1

Operation mode

Refer to parameter [210] on page 31

42008

R/W

2

Capacity

Refer to parameter [212] on page 31

42010

R/W

1

Decimal point

Dec

Description

0

XXXXOO

1

XXXXXO

2

XXXXXX

3

XXXXX.X

4

XXXX.XX

5

XXX.XXX

42011

R/W

1

Increment

Dec

Description

1

X 1

2

X 2

3

X 5

(1) Write this command after writing values to 40031-32 addresses.
(2) Only set point low addresses are used if the set point is programmed as standard.

Programming steps of frequent used ……
Reading a weight value:

1. Read 40003,
2. Check D0=0 and D1=1,
3. If yes, read a weight value ( gross, net or tare),
4. If D0=1, check D0 until system ready,
5. If D1=0, check the error code.


Zero Calibration procedure:

1. Check the low byte of 40033. it should be decimal '1' to start adjustment.
2. Load the decimal '188' to 40030 to start Zero calibration.
3. Check the low byte of 40033. it is decimal '3' during zero calibration process.
4. The low byte of 40033 changes to decimal '1' at the end of the Zero calibration.
5. If the low byte of 40033 is '9', check the high byte of 40033 to understand the calibration error.


Span Calibration procedure:

1. Check the low byte of 40033. it should be decimal '1' to start adjustment.
2. First load the span value to 40031-32 and then load the decimal '220' to 40030 to start Span

calibration.

3. Check the low byte of 40033. it is decimal '4' during span calibration process.