Flintec FT-10xx User Manual

Page 52

Advertising
background image

FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

4. The low byte of 40033 changes to decimal '1' at the end of the Span calibration.
5. If the low byte of 40033 is '9', check the high byte of 40033 to understand the calibration error.


EXPLANATION:

Attention:

For hardware connection details, please refer to the related hardware descriptions in this manual.


Exception codes:

1: Function code is not supported.

2: Out of beginning and ending address range.

3: Invalid value entrance or wrong byte number.

4: Operation error.


Command Examples:
Performing Read and Write operations according (Modbus RTU High-Low) to hex system with the instrument
set to address “0x01”.

Below you will find some command samples;

Description

Hex

Request weight data

01,03,00,00,00,02,C4,0B

Answer of request weight
(weight value is 100000 )

01,03,04,00,01,86,A0,38,4A

Request status data

01,03,00,02,00,01,25,CA

Taring

01,10,00,08,00,01,02,00,02,26,D9

Request tare data

01,03,00,03,00,02,34,0B

Answer of request tare
(tare value is 10000 )

01,03,04,00,00,27,10,E0,0F

Zero Command

01,10,00,08,00,01,02,00,01,66,D8

Request Calibration Status

01,03,00,20,00,01,85,C0

Answer of request Calibration Status
(Instrument is ready for calibration)

01,03,02,00,01,79,84

Zero Calibration

01,10,00,1D,00,01,02,00,BC,A4,6C

Span Calibration Command with Span
value 50000

01,10,00,1D,00,03,06,00,DC,00,00,C3,50,F7,F0

Total LC capacity Command with
Total LC capacity value 100000

01,10,00,1D,00,03,06,00,EC,00,01,86,A0,D4,E0

Average mV/V Command with
Average mV/V value 1.9999

01,10,00,1D,00,03,06,00,FA,00,00,4E,1F,DA,93

Dead load Command with Dead load
value 12345

01,10,00,1D,00,03,06,00,AB,00,00,30,39,87,25

Save the coefficients of eCal
Command

01,10,00,1D,00,01,02,5A,A5,5F,06

Read Voltage of Power Supply value

01,03,00,63,00,01,74,14

Answer of Voltage of Power Supply
(Voltage of Power Supply is 23,5 V)

01,03,02,00,EB,F8,0B

Read digital inputs

01,03,00,C7,00,01,35,F7

Answer of digital inputs
(Input-2 is active)

01,03,02,00,02,39,85

Read digital outputs

01,03,00,C8,00,01,05,F4

Answer of digital outputs
(Output-3 is Active)

01,03,02,00,04,B9,87

Read Setpoint-1 Low

01,03,00,C9,00,02,14,35

Answer of Setpoint-1 Low

01,03,04,00,00,03,E8,FA,8D

Load Set point 1 Low = 5000

01,10,00,C9,00,02,04,00,00,13,88,32,C3