5 calibration block [3, ] scale build, ] calibration – Flintec FT-10xx User Manual

Page 32

Advertising
background image

FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

[21-] Scale Build

The capacity and the resolution of the scale will be defined here.

[210 0] Operation mode

0 : Weighing ( unipolar )
5 : Force

( bipolar )

You will reach the next parameter by pressing the

key.

[212 ] Capacity

Press

key to reach this parameter.


[CAP ] [ XXXXXX ]

Enter the capacity of the scale via

and

keys and confirms the value with pressing

.


[d ] [ XXXXXX ]

Display resolution will be selected by

key and confirmed with

key.

[214 1]

Unit

0 : g

(Gram)

1 : kg

(Kilogram)

2 : t

(Ton)

3 : lb

(Libre)

4 : klb

(Kilolibre)

5 : N

(Newton)

6 : kN

(Kilonewton)


5.5 Calibration Block [3--]

The calibration of the scale will be performed here

after the “Scale Build [21-]” is set .

[30-] Calibration

[300 ] Gravity

This parameter should be used in the scale that will be verified in two stages by gravity adjustment
(in legal Metrologic applications). This parameter should not be touched in other applications.

If you enter a value in this parameter before calibration ( as six decimal digits. Enter 798564 for
9.798564), this value will be assumed as the reference gravity acceleration where the first stage of
the calibration performed. After calibration this parameter will be zeroed. If value of this parameter is
zero, that means no gravity adjustment had been performed after calibration.
In the second stage of verification, the gravity acceleration of the place that the weighing instrument
will be used should be entered (as six decimal digits. Enter 800065 for 9.800065) and exit
programming by saving the changes without entering the calibration (par [ 301 ] ).











Advertising
This manual is related to the following products: