FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

Page 75 of 82

PLC Output to FT-10 CO Input RxPDO 1 (R_UL1)

Bit Number

RxPDO 1 (R_UL1) descriptions

D63 … D56

Set / Reset digital outputs

D55 … D48

Expanded Commands List ( Refer to Table 12.2)

D47 ... D43

Not in use

D42 … D38

Command
List

Bin

Dec

Commands

00000

0

None command is activated

00001

1

Zero

00010

2

Tare

00011

3

Clear

00100

4

Print

00101

5

Adjust zero calibration

Calibration

00110

6

Adjust span calibration

( 1 )

00111

7

Total Load Cell Capacity

( 1 )

eCal
Coefficients

Refer to par.
[313]

01000

8

Average mV/V value

( 1 )

01001

9

Dead Load value

( 1 )

01010

10

Save the coefficients of eCal

01011
01110

11
14

Not used

01111

15

Set Point-1 Low

( 1 )

10000

16

Set Point-1 High

( 1 )

10001

17

Set Point-2 Low

( 1 )

10010

18

Set Point-2 High

( 1 )

10011

19

Set Point-3 Low

( 1 )

10100

20

Set Point-3 High

( 1 )

10101

21

Set Point-4 Low

(1)

10110

22

Set Point-4 High

(1)

10111

23

Set Point-5 Low

(1)

11000

24

Set Point-5 High

(1)

11111

31

Use the Expanded Command list ( Refer to Table 12.2)


D37 ... D33

Read Data
Selection

00000

0

Indicated weight

00001

1

Gross weight

00010

2

Tare weight

00011

3

Calibration Status ( Refer to Table 12.1)

00100
01110

4
14

Not used

01111

15

Set Point-1 Low

( 2 )

10000

16

Set Point-1 High

10001

17

Set Point-2 Low

( 2 )

10010

18

Set Point-2 High

10011

19

Set Point-3 Low

( 2 )

10100

20

Set Point-3 High

10101

21

Set Point-4 Low

( 2 )

10110

22

Set Point-4 High

10111

23

Set Point-5 Low

( 2 )

11000

24

Set Point-5 High

11111

31

Use the Expanded Command list ( Refer to Table 12.2)

D34

New CMD

Toggle

Apply commands which are listed in this table

D33~D0

D33~D37 bits defines the usage of this Dword.


(1) Write this

command with writing values to D0…D33 bits then apply New CMD.

(2) Only set point low addresses are used if the set point is programmed as standard.