FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

40031

R/W

2

Span Calibration Value / LC capacity / mV value / Dead load value

40033

R

1

Calibratio
n Status


D0 .. D7

Calibratio
n Process
Status

De
c

Description

1

Ready for calibration

3

Zero calibration in process …

4

Span calibration in process …

9

Error (Refer to D8 ... D15 )

D8 .. D15

Calibratio
n Errors

1

Calibration Timeout
- Restart calibration

2

ADC Error
- Re-energize the instrument
- If seen again, change the board.

3

Instrument can not be calibrating
- Check load cell cable
- Re-energize the instrument

34

Instrument can not be calibrating
- Load cell signal is very low or too high

35

Calibration Error
- Calibration loading is not enough
- Check test weight loading
- Check load cell connections

36

Calibration load value entry Error
- Test weight is too small. Increase the test
weight

37

Scale unstable
- Wait until scale become stable
- Check grounding wiring

38

The Calibration DIP switch is not ON
position.
- Check the calibration DIP switch.

40034

R/W

37

Not used

40071

R

1

Indicated weight

40072

R

1

Status

40073

R

1

Tare weight

40074

R

1

Gross weight

40100

R

1

Voltage of
Power Supply

Voltage of power supply is indicated with 0.1 V
increment.
For example: 23.4 VDC is indicated as integer 234
value.

40200

R

1

Status of Inputs

D0

Input-1

D1

Input-2

D2

Input-3

D3

Input-4

40201

R/W

1

Status of Outputs

D0

Output-1

D1

Output-2

D2

Output-3

D3

Output-4

D4

Output-5

40202

R/W

2

Setpoint 1 Low

( 2 )

40204

R/W

2

Setpoint 1 High

40206

R/W

2

Setpoint 2 Low

( 2 )

40208

R/W

2

Setpoint 2 High