Er to table, Er to table 9.1, O table 9.1 – Flintec FT-10xx User Manual
Page 58
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014
Expanded Command List (always 32 bit integer)
The “D23 … D16” bits in 2
nd
Dword describes below.
Bit No
Description
D23…D16
Expanded
Commands
List
(continuation
of previous
page)
Bin
De
c
Commands
0000000
0
0
R
Voltage of
Power Supply
Indicated with 0.1 VDC
increment.
0000000
1
0011111
1
1
63
Not in use
0100000
0
64
RW Filter
(1)
Refer to par. [120] on page
29
0100000
1
65
RW Power On Zero
(1)
Refer to par. [202] on page
30
0100001
0
66
RW Zeroing Range
(1)
Refer to par. [203] on page
30
0100001
1
67
RW
Auto Zero
Tracking
(1)
Refer to par. [204] on page
30
0100010
0
68
RW Tare
(1)
Refer to par. [205] on page
30
0100010
1
69
RW
Motion Detector
(1)
Refer to par. [206] on page
30
0100011
0
70
RW Stability Period
(1)
Refer to par. [207] on page
30
0100011
1
71
RW
Operation mode
(1)
Refer to par. [210] on page
31
0100100
0
72
RW Capacity
(1)
Refer to par. [212] on page
31
0100100
1
73
RW Decimal point
(1)
0
XXXXOO
1
XXXXXO
2
XXXXXX
3
XXXXX.X
4
XXXX.XX
5
XXX.XXX
0100101
0
74
RW Increment
(1)
1
X1
2
X2
3
X5
Table 9.1 - Expanded Command List
(1) Write this command after writing values to 1
st
Dword then apply this command with New CMD
Programming steps
of frequent used ……
Reading a weight value:
1.
Check the D12…D15 bits of ‘FT-10 PB Output to PLC Input 2
nd
Dword’.
2. If there is not any error, read a weight value ( gross, net or tare),
Zero Calibration procedure:
1. Check the low byte of Calibration Status. it should be decimal '1' to start adjustment.
2.
Write ‘Adjust Zero Calibration’ command and apply New CMD to start Zero calibration.
3. Check the low byte of Calibration Status. it is decimal '3' during zero calibration process.
4. The low byte of Calibration Status changes to decimal '1' at the end of the Zero calibration.
5. If the low byte of Calibration Status is '9', check the high byte of Calibration Status to understand the
calibration error.
Span Calibration procedure:
1. Check the low byte of Calibration Status. it should be decimal '1' to start adjustment.
2.
Write ‘Adjust Span Calibration’ command after writing test weight values to 1
st
Dword, then apply this
command with New CMD to start Span calibration.
3. Check the low byte of Calibration Status. it is decimal '4' during span calibration process.
4. The low byte of Calibration Status changes to decimal '1' at the end of the Span calibration.
5. If the low byte of Calibration Status is '9', check the high byte of Calibration Status to understand the
calibration error.