FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

[313 ] eCal Calibration

Warning:

The scale capacity and increment shall be entered before performing eCal.


This parameter lets you to perform calibration without using any test weights. FT-10 A/D coefficients
are adjusted in production for increasing eCal accuracy. The calibration coefficients are calculated
by scale capacity, total load cell capacity, load cell full scale output, and estimated dead load. If the
conditions are convenient for zero calibration, you may perform automatic zero adjustment instead

of entering estimated preload. Press

key to start eCal.

[LC.CAP ] [XXXXXX]

Enter total load cell capacity via

and

keys and press

key to go to the next step.

Example : If the weighing system has 4 pcs 1000 kg load cell, enter 4000.

[LC.oUt ] [XXXXXX]

Enter load cell output in mV/V via

and

keys. If the weighing system has more than one

load cell, calculate the mean value of load cells outputs mV/V indicated on the certificates of the load

cells. Press

key to go to the next step.

Example: If load cell outputs are LC1: 2.0010, LC2: 1.9998, LC3:1.9986 and LC4:2.0002, the mean
value will be
Mean of LC outputs = ( 2.0010 + 1.9998 + 1.9986 + 2.0002 ) ÷ 4 = 1.9999 mV/V.

[ZEr.AdJ]
[XXXXXX]
If the scale is empty and you want to make automatic zero adjustment instead of entering estimated

dead load (see next step), press

key. After [ Zero.CA ] appears, press

key for starting

zero adjustment. The display will show [

WAıt ] message during zero adjustment. In this while the

scale must be unloaded and stable. Approximately 10 seconds later the display will prompt you to
save the calibration by [ SAvE ] message below.
If the scale is not empty or you prefer to enter estimated preload value, press the up

key.


[PrE-Ld]
[XXXXXX]

Enter the dead load value of the weighing system in current unit by using

and

keys.

Press the

key to go to the next step.


[ SAvE ]

Save your eCal calibration by pressing

key or press

key to go out without saving your

eCal calibration.

Note:

If you want to make zero adjustment after entering estimated preload value, empty the scale, change

the preload value as (estimated value + display value at empty scale ) or enter parameter [ 310 ] for zero
adjustment.