FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

5.5.1 Analogue Output Block [4--] ( Only FT-10 AN )

The calibration of the analogue output is performed in this sub-block.

[40-] Signal Selection

[400 0] Analogue Output Mode

0 : 4

– 20 mA

1 : 0

– 20 mA

2 : 0

– 5 VDC

3 : 0

– 10 VDC

[401]

Set Analogue Output to calibration range

Press Enter key. [ Ld dEf ] message appears on the display. Press

key for loading default

values or press

key to go next sub block.

[41-] Zero Adjustment

[410]

Coarse Zero Adjustment

The coarse zero adjustment is being performed by pressing

key one after another to increase

the analogue signal level or

key to decrease the analogue signal level.

[411]

Fine Zero Adjustment

The fine zero adjustment is being performed by pressing

key continuously one after another

to increase the analogue signal level or

key to decrease the analogue signal level.

[42-] Span Adjustment

[420]

Coarse Span Adjustment

Coarse span adjustment is being performed by pressing

key continuously one after another to

increase the analogue signal level or

key to decrease the analogue signal level by taking the

full capacity value as reference without placing any weight.

[421]

Fine Span Adjustment

Fine span adjustment is being performed by pressing

key one after another to

increase the analogue signal level or

key to decrease the analogue signal level by

taking the full capacity value as reference without placing any weight