Flintec FT-10xx User Manual

Page 79

Advertising
background image

FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014

Page 78 of 82

01-

RS 485

206

Motion detector

2 = ± 1e

010 Data Format

5 = Modbus RTU
Lo-Hi

207

Stability Period

0.0

011 Baud rate

3 = 9600

21-

Scale Build

013 Address

1

210

Scale Type

0 = Single
Range

014 Data length and Parity 0 = 8 bit, no parity

212

Capacity / d

6000 kg / 1
kg

015 Checksum

0 = Disable

214

Unit

1 = kg

016 Carriage Return

1 = Enable

3--

Calibration Block

017 Line Feed

1 = Enable

30-

Calibration

018 Response Speed

0 = immediately

300

Gravity

03- Ethernet

301

Calibration

030 Ethernet Data Format

0 = No

302

Linearity Correction

031 Ethernet Address

1

31-

Adjustment

032 IP Address

192.168.016.250

310

Zero adjustment

033 Gateway Address

192.168.016.253

311

Span adjustment

034 Subnet Mask

255.255.255.000

312

Span adjustment under load

035 Local Port

502

313

eCal Calibration

036 Response Speed

0 = immediately

4--

Analogue Output Block

04-

Printer

40-

Signal Selection

040 Print out format

2 = Multi line

400

Analogue Output Mode

0 = 4-20mA

041 CN

1 = Will be printed

401

Set Analogue output to calib.
range

042 Minimum print

20

41-

Zero Adjustment

043 Print method

0 = With Print Key

410

Coarse Zero Adjustment

044

Line feed before
printing

00 = 0 F + 0 LF

411

Fine Zero Adjustment

045 Line feed after printing 04 = 0 F + 2 LF

42-

Span Adjustment

046 Form feed

0 = Disable

420

Coarse Span Adjustment

047 Space on the left

3

421

Fine Span Adjustment

048 Quantity of copies

1

70-

Digital Output Functions

05-

Profibus DP

700

Output 1

0 = No

050 Data Format

0 = signed 32 bit

701

Output 2

0 = No

051 Rack Address

0

702

Output 3

0 = No

06-

Profinet

703

Output 4

0 = No

060 Data Format

0 = signed 32 bit

704

Output 5

0 = No

07-

CANopen

8--

Metrological Data Block

070 Data Format

0 = signed 32 bit

80-

Legal Metrology

071 Rack Address

0

800

Calibration counter

1--

Configuration Block

9--

Diagnostic

11-

Start Up

90-

Tests

112 Tare memory

1 = Enable

900

Key Pad testing

113 Auto tare clear

0 = Not used

901

RS 232C testing

115 Key lock

11111

902

RS 485 testing

116 Function key

1 = x10

903

Parallel inputs test

12-

Filter

904

Parallel outputs test

120 Filter

7

905

mV indication

13- Parallel I / 0

99-

Printing Parameters

130 Outputs

0

990

Whole parameters

131 Input 1

0

991

Load default parameters

132 Input 2

0

133 Input 3

0

134 Input 4

0




Advertising