Appendix 1 - unpacking and inspection, Appendix 2 - troubleshooting, Sensor troubleshooting – YSI 5500D-1 User Manual

Page 100

Advertising
background image

YSI 5X00

199

YSI 5X00

198

Appendix 1 - Unpacking and Inspection

Inspect the outside of the shipping carton for damage. If damage is detected, contact

the carrier immediately. Remove the 5X00 from the shipping container. Be care-

ful not to discard any parts or supplies. Confirm that all items on the packing list

are present. Inspect all assemblies and components for damage. Save the original

packing carton. Carriers typically require proof of damage due to mishandling.

The following components are included with the purchase of the 5X00 instrument:

5X00 instrument

006515 Flange Mounting Kit

006506 Desiccant Kit

605877 Instruction Manual

605230 Compression Plugs, 2 each

605229 Compression Plugs, 2 each

605804 Compression Plugs, 4 each

605898 Compression Plugs, 3 each

655384 Dual Male RJ-45 Cable Assembly

655383 Adapter, DB-9 to RJ-45

605880 AC Power Cord, 155 VAC U.S. Plug (AC version only)

605944UL Pluggable Socket Connector for AC Power, 3 pole (AC version only)

506939UL Pluggable Socket Connector for 12 VDC Power, 2 pole

605941UL Pluggable Socket Connector for RS-485, 5 pole

605943UL Pluggable Socket Connector for relays, 12 pole, 2 each

605940UL Pluggable Socket Connector for Auxiliary Inputs, 3 Pole

605942UL Pluggable Socket Connector for DO and AUX inputs, 8 Pole, 2 each

If any parts are damaged or missing, contact YSI or your YSI representative im-

mediately.

Unpacking and Inspection

Troubleshooting

Appendix 2 - Troubleshooting

Sensor Troubleshooting

- When troubleshooting sensors, disable the control/alarm relays.

- Verify 5X00 wiring and configuration.

Issue

possible cause

possible solution

inaccurate or fluctuating

sensor values

improper placement of probe

assembly

See Location Considerations

- Step 2 Locate and Install

Probe Assembly - page

33..

fouled membrane or sensor

cap; damaged membrane or

sensor cap

visual inspection of the

probe assembly required.

Clean membrane or sensor

cap.

improper placement of 5X00 See Location considerations

- Step 1 Install 5X00 - page

27.

bad sensor

bad DO membrane

recalibrate, change mem-

brane, clean sensor, test

sensor outside system

stream in solution of

known value, replace sensor

sensor out of calibration

See Calibration - page

82..

DO elevation configuration

incorrect

Correct DO elevation at

Menus → System → Sensor

Setup→DO→Elevation

menu

DO values are affected by

temperature. If DO values

are incorrect, verify that

the temperature source is

accurate and reporting cor-

rect values or that the user

defined temperature value

is correct for the DO sensor

location. If the temperature

sensor is reporting incorrect

values, the DO mg/L values

are not reliable.

See DO and temperature -

page 155.