Subnet Mask

Menus → System → Communications → Ethernet → Subnet Mask

A Subnet mask address can only be configured when DHCP is disabled. The subnet

mask number helps to define the relationship between the host (computers, rout-

ers, switches, etc.) and the rest of the network. Consult the network administrator

when configuring this menu.

Notes:

- When DHCP is enabled, the Subnet Mask address is not updated at the Subnet

Mask menu.

- AquaManager users should program the Subnet mask either by force serial

port using AquaManager or locally at the 5X00. Poll node using RS232 com-

munication . Next upload configuration information including: static IP, port,

subnet, and gateway addresses.

- Numbers between 0-255 are valid for each of the four (4) Subnet Mask IP

Address fields.

Gateway

Menus → System → Communications → Ethernet → Gateway

The Gateway address can only be configured when DHCP is disabled. A gateway IP

refers to a device on a network which sends local network traffic to other networks.

Consult the network administrator when configuring this menu.

Notes:

- When DHCP is enabled, the Gateway address is not updated at the Gateway

menu.

- AquaManager users should program the Gateway address either by force

serial port using AquaManager or locally at the 5X00. Poll node using RS232

communication . Next upload configuration information including: static IP,

port, subnet, and gateway addresses.

- Numbers between 0-255 are valid for each of the four (4) Gateway IP Ad-

dress fields.

Port

Menus → System → Communications → Ethernet → Port

The default port number for the ethernet module is 10001. Consult the network

administrator when configuring this menu.

Note:

- AquaManager users should program the port number either by force serial

Configuring the 5X00

port using AquaManager or locally at the 5X00. Poll node using RS232 com-

munication . Next upload configuration information including: static IP, port,

subnet, and gateway addresses.

DNS Address

Menus → System → Communications → Ethernet → DNS

Domain Name Service (DNS) refers to the server, or workstation, used to convert

domain names into IP addresses.

Notes:

- With DHCP enabled, the serial to ethernet device will acquire SMTP and DNS

addresses during initialization.

- When DHCP is enabled, the DNS address is not updated at the IP address

menu.

- Numbers between 0-255 are valid for each of the four (4) DNS IP Address fields.

Email Addresses

Menus → System → Communications → Ethernet → Email Address

Email

Address

Email

Address

CC/Email

Address

Up to four email addresses can be configured for 5X00 emails. At a minimum, the

first email address and the CC/email address must be configured. If the CC/email

address is not configured, 5X00 email alarms will contain no alarm information.

The email ‘from’ address will be the email address entered in the first Email Ad-

dress menu.

Notes:

- See Send Test Email Msg - page 101 to verify proper configuration of ethernet

menus.

- On a network, both master and slave alarm emails are sent to the configured

email addresses in the master.

- Test the alarm email delivery by creating an alarm condition at the 5X00 and

verify the email(s) is received.

- The CC/Email address must be configured when using 5X00 email alarming.

Specific alarm system information will not be sent if the CC/Email address

menu is not configured with a valid email address.

- Use AquaManager to easily configure email addresses using the PC keyboard.

See AquaManager on line help for additional information.

- If Autopoller is being used for AquaManager email alarming on a TCP/IP

connection, it is strongly recommended that the 5X00 email alarming using

the serial to ethernet device be disabled. When the ethernet device sends

email alarms, all Autopoller SMS communication is suspended. This could

Configuring the 5X00