Location considerations – YSI 5500D-1 User Manual

Page 18

Advertising
background image

YSI 5X00

35

YSI 5X00

34

For accurate DO measurements, ensure that the temperature reading for each DO

channel is representative of the temperature at that DO sensor’s location.

This is important when using the same temperature input source for multiple DO

sensors. Example - if the temperature source for DO1 and DO2 is the temperature

sensor connected to Aux 3 at the DO1 sensor location, it is important that the tem-

perature at the DO2 sensor location is the same as at DO1 sensor location.

See Chapter 4 Sensor Setup - page 137, for information on configuring 5X00

sensor systems. User defined sensor configuration items include:

- Display format (i.e. units of measure. For example, mg/L or % saturation,

°C or °F)

- Temperature source

- Salinity input value (user entered salinity value of the water being moni-

tored)

- Elevation monitoring system

- Setpoint, control, and alarm values and ranges

- Control and alarm relay assignments

Location Considerations

Readings gathered at the sensor location(s) are data logged and processed by the

5X00 instrument. Output control and alarm devices are configured to respond to

sensor values. Therefore, it is essential to have accurate readings at the sensor loca-

tion. Locate sensors according to the below specifications.

Locate sensor end of probe assembly:

• in a location that represents the entire system

• where the water flow is at least 3 inches per second (7.6 cm/sec) if using the

2.0 mil PE membrane or at least 6 inches per second (15.24 cm/sec) for 2.0

mil Teflon membrane

• in a location that is free from debris (e.g. algae); ideally at a 45˚ angle from

vertical position and directed with the flow - figure 3.10

5400

5400

Figure 3.10

• in a well mixed, free-flowing area; midstream and mid-depth typically gives

representative readings; the probe assembly and cable are fully submersible

up to the cable connector

submersed at all times so liquid level fluctuations do not expose the sensors

to the atmosphere

so the sensor/probe cabling is not routed near rotating machinery and/or

equipment involving electrical switching or regulation

• so that the sensor assembly and cabling are located away from sources of

electrical interference such as UV sterilizers, florescent lighting, ballasts,

pumps, any high voltage peripheral devices, etc.

CAUTION: run high and low voltage cabling through separate bulkhead and

conduit

where regular maintenance, including calibration can easily be done. Cali-

brations are performed using the 5X00 front panel.

Note:

- The YSI 5422 DO sensor is field replaceable and can be replaced without

replacing the probe cabling. The 5420 and 5421 DO sensors are not replace-

able. If a 5420 or 5421 sensor requires replacement, the entire probe/cable

assembly needs to be replaced.

Locate sensor end of probe assembly:

• in a location that represents the entire system

• Insert the probe into sample. Move the probe in the sample to release any

air bubbles and to provide a fresh sample to the sensor cap. This movement

is only necessary when first inserting the probe into the sample.

5500

Installation and Wiring

Installation and Wiring