AquaManager and the System Menu

AquaManager users can configure most 5X00 system menus from a PC. After polling

the 5X00, use “Systems Management” from the node window to view and upload

5X00 configuration. System menus are configured at the Setup tab selections.

Notes:

- Always manually poll the 5X00 after uploading configuration changes and

verify uploaded information is retained. When configuration changes are

uploaded using AquaManager at the same time configuration changes are

being made by a user in the menus on a 5X00 instrument, the AquaManager

uploads may not be saved depending on the timing of keystrokes at the 5X00.

- The Run Screen must be displayed during normal operation. If the 5X00

is not at the run screen during AquaManager uploading, pressing

at

the 5X00 will only save certain configuration changes.

-

See Chapter 5 AquaManager - page 175 and AquaManager online help for ad-

ditional information.

Sensor Setup

Menus → Sensor Setup → Aux 1-6

→ DO 1-4

Aux, DO, and temperature system set points, control (set point or PID/PWM),

alarm, and sensor system relays are configured in the Sensor Setup menus. Eight

onboard relays can be configured for control and alarm output devices. Enabled

relays energize when a control or alarm system is active. Alarm notifications can be

sent via email or cell phone (SMS messaging) - page 91.

Notes:

- For normal operation, the 5X00 must be at the Run Screen. Alarm func-

tionality (email alarms, activation of assigned alarm relays, alarm icons,

and buzzer) is suspended when 5X00 is not at Run Screen.

- Sensor system must be enabled in order to display values at Run Screen.

Control and/or alarm relays will not energize regardless of value if the

sensor system is disabled.

- There is a 30 second delay at power up before control, timer, and alarm systems

become active.

- There is a 30 second sensor system hold off when sensor configuration is

changed. See 30 Second Sensor System Hold Off - page 146.

- Calibrate sensor prior to use. See Calibration - page 83.

- “Over” and “Under” readings are displayed when DO sensor is measuring

outside it’s operating range - see DO range specification - page 14 DO

control and alarm functions including relays are suspended when a sensor is

reading “Over” or “Under.”

- “Over” and “Under” are not displayed for aux temperature systems when tem-

perature values are outside the temperature operating range - see Temperature

Range Specification - page 15. Temperature system controls and alarms will not

operate correctly when values are outside the temperature operating range.

- DO “Over” and “Under” readings most likely indicate that the sensor needs

to be serviced or replaced, or a bad probe or cable connection exists.